Specifications per Omega LCGD/LCMGD datasheet (sheet F-27/F-28) and instruction sheet M0443/0707.
Best For
- Compression force, weight and reaction-load measurement on flat-mounted fixtures
- Press, clamp and crimp monitoring on test stands and production equipment
- R&D and QC test rigs needing a NIST-traceable, low-profile sensor
- OEM machine integration where space is tight and stability matters
Warranty
Backed by Omega's 1-year limited warranty (13 months from date of purchase, including a 1-month shipping grace period) against defects in materials and workmanship.
Key Features
All Stainless Steel Construction
Fully stainless body and integral spherical load button for long-term stability and reliability in severe industrial environments.
Integral Spherical Load Button
A machined load button is built into the cell, giving a self-aligning point-contact interface for clean compression-only loading.
Low-Profile Miniature Form
At 1.50" outer diameter and 0.63" high, the cell drops into tight assemblies and mounts directly on any flat surface at least 1.50" across.
5-Point NIST-Traceable Calibration
Ships with a 0-50-100-50-0% NIST-traceable calibration certificate and 59 kΩ shunt-calibration data for in-field verification.
High Static Accuracy
±0.25% FSO linearity, ±0.20% FSO hysteresis and ±0.10% FSO repeatability across the rated range.
Sealed and Overload Protected
IP65 sealing, a 350 Ω bridge, a shielded 5 ft PTFE cable, and 150% safe / 300% ultimate overload margins.
Technical Specifications
| Measurement Type | Compression only |
| Rated Capacity | 10,000 lbf (44,484 N) |
| Rated Output | 2 mV/V nominal (see calibration sheet for exact output) |
| Excitation | 10 Vdc |
| Linearity | ±0.25% FSO (BFSL) |
| Hysteresis | ±0.20% FSO |
| Repeatability | ±0.10% FSO |
| Zero Balance | ±2% FSO |
| Bridge Resistance | 350 Ω minimum |
| Deflection | 0.076 mm (0.003") at full scale |
| Safe Overload | 150% of capacity |
| Ultimate Overload | 300% of capacity |
| Calibration | 5-point NIST-traceable; 59 kΩ shunt data included |
| Compensated Temp Range | 16 to 71°C (60 to 160°F) |
| Operating Temp Range | -54 to 121°C (-65 to 250°F) |
| Thermal Effect on Zero | ±0.009% FSO/°C |
| Thermal Effect on Span | ±0.018% rdg/°C |
| Protection Class | IP65 |
| Construction | Stainless steel |
| Outer Diameter (D1) | 1.50" |
| Height (H) | 0.63" |
| Mounting | 3 holes, #6-32 UNC x 0.25" deep on 1.25" bolt circle; flat-surface mount |
| Electrical Connection | 1.5 m (5 ft) shielded PTFE cable, pigtail end |
| Cable Code | Red +Exc, Black -Exc, White +Sig, Green -Sig |
| Compatible Meters | DP25B-S, DP41-S and DPiS |
Documentation
Related Products
Frequently Asked Questions
- What does the Omega LCGD-10K measure?
- It is a compression-only miniature load cell rated for 10,000 lbf (44,484 N) full scale. It reads force applied through its spherical load button and is not intended for tension.
- What output and excitation does it use?
- Rated output is 2 mV/V nominal at 10 Vdc excitation; the exact factory output is printed on the supplied calibration sheet.
- Does it come calibrated?
- Yes. Every cell ships with a 5-point (0-50-100-50-0%) NIST-traceable calibration certificate and 59 kΩ shunt-calibration data.
- How is the LCGD-10K mounted?
- It rests on a flat surface at least 1.50" in diameter and is secured with three #6-32 UNC screws on a 1.25" bolt circle. Load is applied to the integral spherical button.
- What is the overload rating?
- Safe overload is 150% of rated capacity; ultimate overload is 300%. Operating beyond the safe limit can shift calibration.
- Which meters work with it?
- It pairs with the Omega DP25B-S, DP41-S and DPiS digital panel meters, which supply bridge excitation and scale the mV/V signal to engineering units.
1-Year Limited Warranty. Omega warrants this load cell free of defects in materials and workmanship for 13 months from date of purchase (a 1-year warranty plus a 1-month shipping grace period). Coverage excludes damage from misuse, overload beyond rated limits, improper interfacing, or unauthorized modification.
- Load Cells:
- Industrial Miniature
- Measurement Type:
- Compression
- Capacity:
- 10000 lbf
- Rated Output:
- 2 mV/V
- Accuracy:
- ±0.25% FSO
- Excitation:
- 10 Vdc
- Construction:
- Stainless Steel
- Protection Rating:
- IP65
- Outer Diameter:
- 1.50 in
- Cable Length:
- 5 ft
- Calibration:
- NIST-Traceable