The AimSafety PM400-STN is a bump test and calibration station that tests, calibrates, configures and updates firmware on up to four PM400 four-gas monitors at a time without a computer.
Specifications on this page are taken from the Macurco AimSafety product data workbook, the PM400 Series Data Sheet and the AimSafety line card.
Best For
Key Features
Technical Specifications
| Functions Performed | Bump test and calibration |
|---|---|
| Capacity | Up to four PM400 monitors simultaneously |
| Compatible Monitors | AimSafety PM400-IR and PM400-P |
| Configuration | Automatic update of PM400 configuration and firmware |
| Logging | Event logging through PC or PM Link |
| Battery | Rechargeable, approximately 1000 bump tests per charge |
|---|---|
| Computer Required | No, for normal operation |
| Size | 20.6 in. x 16.8 in. x 8.1 in. (52.4 x 42.8 x 20.6 cm) |
|---|---|
| Weight | 23 lb. (10.4 kg) |
| Color | Black |
| Operating Temperature | 41 to 104 °F (5 to 40 °C) |
| Country of Origin | South Korea |
| ECCN | EAR99 |
Documentation
Related Products
Frequently Asked Questions
Which monitors does the AimSafety PM400-STN work with?
The PM400-STN is the bump test and calibration station for the PM400 series, covering both the PM400-IR infrared and PM400-P catalytic bead models. PM100 single-gas monitors use the separate PM100-STN.
How many PM400 monitors can the station handle at once?
Four. The station bump tests and calibrates up to four PM400 multi-gas units simultaneously.
Does the PM400-STN update firmware?
Yes. It automatically updates PM400 configuration and firmware on the monitors it docks, so a fleet stays on a single revision without individual PC sessions.
Which calibration gas does the PM400-STN use?
The PM400 quad mixture cylinder, part 37-0551-3344-2, supplying CO 100 ppm, H2S 25 ppm, CH4 50 % LEL and O2 18 % by volume in a 58L cylinder.
Does the PM400-STN need a computer?
Not for normal operation. A computer or the PM Link interface is used for event log retrieval and configuration changes.
- Product Type:
- Bump Test and Calibration Station
- Compatible Series:
- PM400
- Capacity:
- 4 Monitors
- Compatible Gases:
- CO / H2S / O2 / LEL
- Power Source:
- Rechargeable Battery
- Computer Required:
- No